CALIBRATION CERTIFICATE
Customer Details:
Certificate No:
Received Date:
Calibration Date:
Recom Due Date:
Cert. Issued Date:
Calibrated at:
The piston operated device calibration carried out with respect to standard as per ISO 8655-6:2002 in compliance with National/International Standards.
Unit Under Calibration:
Description Serial No
Make ID no
Model Calibration Points(µl)
Volume Range(µl) Condition on Receipt
Least Count(µl) Accuracy(µl)±
Location No of Channel
Environmental conditions:
Temperature(°C) Relative Humidity(%) Air Pressure(hPa)
Standard's Traceability Details: ⚙ Standards & Traceability Master
Sl.No Equipment / Reference Standard Sl No Calibrating Agency Traceability No Valid upto Cert
1
📄
2
📄
Calibration Results (Mechanical Volumes):
Sl No Device Readings(µl) Calculated Mean Volume at 27 °C(µl) Error(µl) Uncertainty(±µl)
1 100 100.13 0.13 0.29
2 500 499.31 -0.69 0.75
3 1000 1000.66 0.66 1.23
As per ISO 8655-2:2002 systematic error, Random error and Error limits:
Sl.No Device Readings(µl) Calculated Mean Volume at 27 °C(µl) Systematic Error(µl) Syst Error Limits(±µl) Random Error(µl) Random Error Limits(±µl)
1 100 100.13 0.13 0.80 0.13 0.30
2 500 499.31 -0.69 4.00 1.11 1.50
3 1000 1000.66 0.66 8.00 1.90 3.00
Remarks:
* Calculated Mean Value given is the average of 10 measurements
* Equipment used for calibration were calibrated & Traceable to National and International Standard
* The reported Expanded Uncertainty is calculated at 95.45 % C.L. with coverage factor k=2
* This calibration certificate is valid for scientific & industrial purpose only
* Ref PSRF No:
* This report refers only to particular item submitted for calibration
Status: PASS (According to ISO 8655-2:2002 permissible limits)
Calibrated by
Calibration Engineer
Page No 1 of 2
Authorised Signatory by
QM / TM
Annexure to Certificate No: 801
Calibration Date: --
CALIBRATION REPORT - RAW GRAVIMETRIC OBSERVATIONS (ANNEXURE)
Individual 10 Weighing Observations per Volume Target as per ISO 8655-6:2002 / ISO 8655:2022 Gravimetric Procedure
Customer: --
Instrument: --
Serial No: --
Range: --
Standard Balance: --
Z-Factor: µL/mg
Water Temp:
1. Minimum Volume: µL (Vmin)
Run Mass mi (mg) Volume Vi (µL)
1 100.02
2 100.00
3 99.97
4 100.03
5 100.01
6 99.99
7 100.04
8 99.98
9 100.02
10 100.00
Target Volume Vs:100.00 µL
Mean Mass m̄:0.000 mg
Mean Volume V̄:0.00 µL
Systematic Error es:0.00 µL
Inaccuracy A%:0.00%
ISO Permissible Limit ±A%:±%
Standard Deviation s:0.00 µL
Random Error CV%:0.00%
ISO Permissible Limit CV%:≤%
Uncertainty U (k=2):±0.00 µL
Result:PASS
2. Middle Volume: µL (Nominal/2)
Run Mass mi (mg) Volume Vi (µL)
1 500.05
2 500.00
3 499.94
4 500.08
5 500.02
6 499.97
7 500.03
8 499.98
9 500.02
10 500.01
Target Volume Vs:500.00 µL
Mean Mass m̄:0.000 mg
Mean Volume V̄:0.00 µL
Systematic Error es:0.00 µL
Inaccuracy A%:0.00%
ISO Permissible Limit ±A%:±%
Standard Deviation s:0.00 µL
Random Error CV%:0.00%
ISO Permissible Limit CV%:≤%
Uncertainty U (k=2):±0.00 µL
Result:PASS
3. Nominal Volume: µL (Nominal)
Run Mass mi (mg) Volume Vi (µL)
1 1000.09
2 999.99
3 999.94
4 1000.04
5 1000.01
6 999.97
7 1000.03
8 999.98
9 1000.02
10 1000.00
Target Volume Vs:1000.00 µL
Mean Mass m̄:0.000 mg
Mean Volume V̄:0.00 µL
Systematic Error es:0.00 µL
Inaccuracy A%:0.00%
ISO Permissible Limit ±A%:±%
Standard Deviation s:0.00 µL
Random Error CV%:0.00%
ISO Permissible Limit CV%:≤%
Uncertainty U (k=2):±0.00 µL
Result:PASS
* Gravimetric Conversion: Vi = mi × Z (Z-Factor temperature & pressure correction per ISO 8655-6).
* Systematic Error (Inaccuracy): es = V̄ - Vs; A% = (es / Vs) × 100%.
* Random Error (Imprecision): s = √[Σ(Vi - V̄)² / (n - 1)]; CV% = (s / V̄) × 100%.
* Expanded Uncertainty: U = k × √[(s / √n)² + ubal²], with coverage factor k=2 (approx. 95.45% confidence interval).
Observed & Recorded by
Calibration Engineer
Calibration Engineer
Page No 2 of 2
Verified & Approved by
Quality Manager / Technical Manager